The speech processor also displays the signal level
picked up by the microphone in a 12 segment
meter. The meter fluctuates according to the level
of sound received, with 12 being the highest level
and 0 the lowest level.
To adjust microphone sensitivity:
Press the Select ( ) button until the ‘M’ symbol
appears on the LCD panel.
To increase the sensitivity, press the Up ( )
button.
To decrease the sensitivity, press the Down ( )
button.
If you need to set the microphone sensitivity
consistently higher than ten, consult your clinician.
The microphone sensitivity setting is saved with
the current program. If you select another program,
the microphone sensitivity will change to the
setting saved with that program.
Using Autosensitivity
When autosensitivity is turned on, the microphone
sensitivity will decrease automatically in a noisy
environment. To activate autosensitivity press the
autosensitivity button ( ). When autosensitivity is
on, the speech processor displays the letter 'A' on
the LCD panel.
Autosensitivity on
SPrint User Manual 29